LS1

LS1

LS1

  • Calibration with ambient air
  • Linear measurement signal
  • For applications up to 1400 °C
  • Fast reaction and response
  • Increased measurement accuracy in the range 0 ... 21 % O2
FeatureValue

Measuring system

Zirconium dioxide cell / current probe

 

Measuring principle

Current probe (Current mode)
Probe current = linear function ƒ(O2 content).

Probe current for 21 Vol.% O2

300 to 600 mA

Measuring cell operating temperature (sensor)

750 to 850 °C

Warm-up time

60 min. to standby mode at 21 % O2
120 min. to thermal equilibrium

Adjustment time after measurement value jump

t90 < 20 sec with standard metering gas extractor, 450 mm long

Measuring range

0-21 Vol. % O2 calibration with air

Measuring accuracy

± 0,2 Vol. % O2 after calibration with
Lambda Transmitter LT1

Type of protection

IP42

Probe signal

0 ... 400/600 mA

TUV suitability approval

 

13 and 17.BlmSchV with Lambda Transmitter LT1

Product approved by: